4-3
Front Panel
Ch1 & Ch2 Outputs
The Channel 1 and Channel 2 outputs can be con-
figured to output a voltage from -10 V to +10 V
proportional to X, Y, R, θ, or Traces 1 through 4.
±10 V is full scale. The outputs can source 10 mA
maximum.
Signal Inputs
The input mode may be single-ended, A, or diffe-
rential, A-B. The A and B inputs are voltage inputs
with 10 M, 25 pF input impedance. Their connec-
tor shields are isolated from the chassis by 10
(Ground) or 1 k (Float). Do not apply more than
50 V to either input. The shields should never
exceed 1 V. The I (current) input is 1 k to a virtu-
al ground.